DPG5000
PRESSURE GAUGES
M3366/0606
User’s Guide
http://www.omega.com
Vue de la page 0
1 2 3 4 5 6 ... 11 12

Résumé du contenu

Page 1 - User’s Guide

DPG5000PRESSURE GAUGESM3366/0606User’s Guidehttp://www.omega.come-mail: [email protected]

Page 2 - OMEGA...Of Course!

DPG5000 M3366/0606Notes8

Page 3 - Table of Contents

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology

Page 4 - • Warranty/Disclaimer

OMEGAnet® On-Line Service Internet e-mail www.omega.com [email protected] North America:USA: One Omega Drive, P.O. Box 4047 ISO 9001 Certi

Page 5 - M3366/0606 DPG5000

Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course!TEMPERATURE✓ Thermocouple, RTD & Thermistor Probes, Conn

Page 6

Table of ContentsWarranty/Disclaimer ...2Return Requests/Inquiries ...

Page 7

• Warranty/DisclaimerOMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date

Page 8 - Typical 4-20 mA Circuit

• Return Requests/InquiriesDirect all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S)

Page 9 - External Power Plug

• InstructionsNOTE: All units are factory calibrated prior to shipment.1. Zero TrimmingNOTE: Some units are equipped with zero “knob”.If it becomes ne

Page 10 - DPG5000 M3366/0606

4. Re-CalibrationThis procedure requires a known pressure source of at least ±0.1% accuracy in order to fully utilize the accuracy potential of the DP

Page 11 - OMEGA Products

7. 4-20 mA Transmitter OptionWhen equipped with this option the gauge no longer operates from batteries but instead is “Loop Powered”. The loop connec

Page 12 - OMEGAnet

AFBECD6 Pin Connector8. External Power (24 VDC) and 0 to 5 VDC Analog Output OptionWhen equipped with this option the gauge no longer operates from ba